4 Operating in EI Mode
To Set Monitors for MSD Temperature and Vacuum Status
5977C GC/MSD Operating Manual 93
5 To set a monitor’s alarm, double-click a monitor displayed in the Instrument
Control view to open that monitor’s dialog for setting alarms. (See Figure 27.)
a Select the Set Alarm check box.
b Set the Warning Level, Alarm Level, and Below Minimum to appropriate
values.
c Enter descriptive text in the Monitor Label field if the default label is not
appropriate.
d Click OK to finish the monitor’s alarm configuration.
6 To make the new settings part of the method, save the Method.
Figure 27 Monitor’s alarm settings