34 Agilent Gas Sample Selector for 8860 and 8890 Safety and Installation Manual
3 Installation
Site preparation
Site preparation
Leave enough space on the left side of the GC to place the bag version selector
(40 cm).
Installation requires 1 power outlet.
Other materials required (not provided):
1/8-inch tubing, terminating in a 1/8-inch male Swagelok fitting, that is long
enough to connect the Selector sample exhaust vent to the laboratory exhaust
system or fume hood.
Compressed air supply, terminating in a 1/8-inch male Swagelok connector
(for valve actuation) at 380 kPa (55 psi).
Purge gas supply, nitrogen, air, or helium, terminating in a 1/8-inch male
Swagelok connector.
Figure 13 Shipping kit contents for the bag version Gas Sample Selector.
1
3
2
4
7
6
8
5