16 G7200-90210 (Rev. A) MassHunter Workstation Quick Start Guide
9 Set the Integration time per Point for each mass:
a Click on the row for the mass in the column for the appropriate step (i.e.
Step 1 for Hydrogen mode, Step 2 for Helium mode) as specified in the
Reaction Gas column in the Table 4 on page 18. You can select
multiple masses by <Ctrl>+clicking on them.
b Enter the Integration time per Point as specified in Table 4. The
integration times in the table are generally relatively short, and can
be extended as necessary.
c Click the Enter button. The per Point integration time and calculated
per Mass integration time will be updated in the table.
d Repeat for each mass in the table. For masses such as 71 that are
monitored in both steps, remember to enter integration times for both.
9a