Maintenance and Spare Parts
Cary Eclipse Fluorescence Spectrophotometer User’s Guide 39
Checking
To check a fuse:
1 Disconnect the Cary Eclipse instrument from the mains power
supply.
2 Open the hinged cap and pull out the fuse holder (lower red
section). Both fuses are held in this fuse holder.
3 Check that each fuse is the correct type and that they are not
damaged. If necessary, replace the fuse (refer to Page 38).
4 Place the fuses into the red fuse holder and push this red fuse
holder into the main fuse holder and then lower the cap.
5 Reconnect the Cary Eclipse instrument to the mains power
supply and turn on the instrument.
NOTE
If a fuse repeatedly blows, it may indicate other problems with the Cary Eclipse
instrument.