EasyManua.ls Logo

Agilent Technologies E4404B - Page 67

Agilent Technologies E4404B
78 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
5-6 Chapter 5
Making GSM Cable and Antenna Measurements
Making Cable Fault Location Measurements
Table 5-1 Windowing Functions Available for the Cable Fault Location
Measurement
Function Description Max
side-lobe
level
Side-lobe
roll off
Max main
Rectangular The function results in no
windowing.
–13.261 dB 20 dB/decade –3.9224d
Flat Top The default value. A five term
flat top window. A good
window to use when making
amplitude measurements of
relatively pure tones.
–95.1 dB +/– 0.0020
Gaussian A five term cosine window
which resembles a Gaussian
window.
–125.4 dB –0.680056
Hanning A window which has good
frequency resolution and
reasonably good side lobe-lobe
roll-off, but poor main-lobe
flatness and relatively large
side-lobe peaks.
–31.46730784 60 dB/decade –1.423622

Table of Contents