Measurement Uncertainty Analysis - Power Reference Level Test C
E4416A/E4417A Power Meters Service Guide C-143
Sensor Calibration Factor Drift
The Agilent Standards Laboratory at South Queensferry has determined the
following average sensor drift value.
Average Sensor Drift < 0.06% per year
Therefore,
U(Drift) = 0.0006
Resistance
The uncertainty of the resistance measurement on the 3458A DMM is given
from:
Error = (ppm of reading + ppm of range) + (temperature coefficient) + (factory
traceability)
with: -
Conditions: 1 Year Cal and temperature variance +/- 5 °C without ACAL
Reference: Agilent 3458A Operating Manual Appendix A
For 1 kΩ scale the specified accuracy is;
10 ppm of reading + 0.5 ppm of range
with: -
1 ppm of Reading + 0.1 ppm of Range per
o
C (temp coefficient Tcal ±5
o
C)
and, 3 ppm of Reading (for factory traceability)
NOTE
This value is only valid for 8478B sensors calibrated to 0.2% accuracy.
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com

Table of Contents

Other manuals for Agilent Technologies E4416A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies E4416A and is the answer not in the manual?

Agilent Technologies E4416A Specifications

General IconGeneral
BrandAgilent Technologies
ModelE4416A
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals