1 Specifications
10 E4416A/E4417A Power Meters Service Guide
Sampling Characteristics
Rear Panel Inputs/Outputs
Sampling Rate: 20 Msamples/second.
Sampling Technique: continuous sampling.
Recorder Output(s): Analog 0 to 1 V, 1 kΩ output impedance, BNC
connector (Two outputs are available on the
E4417A: Channel A and B).
Remote
Input/Output:
TTL Output: used to signal when measurement has
exceeded a defined limit.
TTL Input: initiates zero and calibration routine
Connector type: RJ- 45 Series shielded modular
jack assembly.
TTL Output: high=4.8 V max, low=0.2 V max.
TTL Input: high=3.5 V min., 5 V max;
low=1 V max, 0.3 V min.
RS- 232/422
Interface:
Serial interface for communication with an
external controller, Male/Plug 9 pin
D- subminiature connector.
Trigger In: Accepts a TTL signal for initiating measurements,
BNC connector.
Trigger Out Outputs a TTL signal for synchronizing with
external equipment, BNC connector.
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com

Table of Contents

Other manuals for Agilent Technologies E4416A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies E4416A and is the answer not in the manual?

Agilent Technologies E4416A Specifications

General IconGeneral
BrandAgilent Technologies
ModelE4416A
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals