100 Chapter 4
Replacement Procedure
Handle Assembly Replacement
Handle Assembly Replacement
Tools required
torque screwdriver T20
Removal Procedure
Refer to Figure 4-25 for this procedure.
Step 1. Remove the four TORX T20 screws (item 1) fastening the Handle Assembly.
Step 2. Remove the end caps (both side) (item 2) from the Handle Assembly.
Step 3. Detach parts at both sides (item 3) of the Handle Assembly.
Step 4. Remove the strap cover (item 4) of the Handle Assembly.
Replacement Procedure
Step 1. Reverse the order of the removal procedure.