Chapter 2 33
Troubleshooting
Troubleshooting Using Diagnostic Test
2. Troubleshooting
Figure 2-15 Test dialog box
To execute the test, click “Execute” button. To return the E5061B
Diagnostic test program dialog box (Main Menu), click “Back to Main
Menu”.
The test program will prompt you to connect cables or Short
termination to the E5061B. Follow the instructions as shown below on
the E5061B display for performing the test.
Figure 2-16 Setup dialog box
The test result file named “resultDT.txt” is created on the drive D
(D:\Agilent\Service\Log\) of the E5061B after “Exit” button of the
Main Menu is pressed. The text file can be read and edited with a PC.
Diagnostic Test Failure Troubleshooting
Table 2-3 and Table 2-4 represents the contents of the diagnostic tests and the relationships
of failed tests to probable faulty board assemblies. If the instrument fails the diagnostic