66 Chapter 4
Replacement Procedure
A60 CPU Module Replacement
A60 CPU Module Replacement
Tools Required
torque screwdriver, T15, and T20
torque screwdriver, T10 (set to 9 in-lb)
Removal Procedure
Refer to Figure 4-7 for this procedure.
Step 1. Remove the Outer Cover as described in “Outer Cover Replacement” on page 54.
Step 2. Remove the guide bar as described in “Power Supply Assembly Replacement” on page 62.
Step 3. Remove the Hard Drive as described in “Hard Disk Assembly Replacement (Option 020)”
on page 60.
Step 4. Disconnect the three cables (item 1) on the A60 CPU Module.
Step 5. Remove the two S-ATA cables and power supply cable (item 2) from the A60 CPU
Module.
Step 6. Remove the three TORX T10 screws (item 3) fastening the Chassis Assembly for the A60
CPU Module with Hard Disk Assembly.
Step 7. Remove the two TORX T10 screws (item 4) fastening rear panel of the Chassis Assembly
for the A60 CPU Module with Hard Disk Assembly.
Step 8. Lift the A60 CPU Module with Hard Disk Assembly.
Step 9. Remove the two TORX T10 screws (item 5) fastening the A60 CPU Module for the Hard
Disk Assembly.
Replacement Procedure
Step 1. Reverse the order of the removal procedure.
NOTE Set a torque to 9 in-lb for torque driver when fasten the TORX T10 screws (item 6, item 7
and item 8).

Table of Contents

Other manuals for Agilent Technologies E5061B

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies E5061B and is the answer not in the manual?

Agilent Technologies E5061B Specifications

General IconGeneral
BrandAgilent Technologies
ModelE5061B
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals