68 Chapter 4
Replacement Procedure
24 Bit I/O Assembly Replacement
24 Bit I/O Assembly Replacement
Tools Required
torque screwdriver, T15, and T20
torque screwdriver, T10 (set to 9 in-lb)
flat edge screwdriver
Removal Procedure
Refer to Figure 4-8 for this procedure.
Step 1. Remove the Outer Cover as described in “Outer Cover Replacement” on page 54.
Step 2. Remove the Front Panel as described in “Front Panel Replacement” on page 56.
Step 3. Disconnect the flat cable (item 1) from the 24 Bit I/O Assembly and the DSP Assembly.
Step 4. Release the flat cable from the clamp (item 2).
Step 5. Disconnect the cable (item 3) from the 24 Bit I/O Assembly and the A50 CPU Module.
Step 6. Release the cable from the clamp (item 4).
Step 7. Remove the two TORX T10 screws (item 5) fastening the 24 Bit I/O Assembly.
Step 8. Slide forward the 24 Bit I/O Assembly.
Replacement Procedure
Step 1. Reverse the order of the removal procedure.
NOTE Set a torque to 9 in-lb for torque driver when fasten the TORX T10 screws (item 5).

Table of Contents

Other manuals for Agilent Technologies E5061B

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies E5061B and is the answer not in the manual?

Agilent Technologies E5061B Specifications

General IconGeneral
BrandAgilent Technologies
ModelE5061B
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals