EasyManua.ls Logo

Agilent Technologies E5061B - Page 36

Agilent Technologies E5061B
123 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Table 2-4 Diagnostic tests failure troubleshooting information (option
115/117/135/137/215/217/235/237)
Test
No.
Test group Failured Test Probable faulty board assembly
A11 Source
Module
A3/A4 Receiver
Module
1 DC-BUS Test ALL ###
A11 Source
Module
### #
A3/A4 Receiver
Module
###
2 Internal Level Monitor Test ALL ## ###
A11 Source
Module
### #
3 RF Output Level Range Test ### #
4 RF Output Level Power Sweep
Test
### #
5 Receiver Absolute Measurement
Test
# ###
7 Receiver IF Ranging Test # ###
Chapter 2 35
Troubleshooting
Troubleshooting Using Diagnostic Test
2. Troubleshooting
###: Most suspicious assembly
##: Suspicious assembly
#: Possible faulty assembly
NOTE Diagnostic tests 1 and 2 are common to the power on self test.