3-38 Service Guide E8364-90026
Tests and Adjustments PNA Series Microwave Network Analyzers
Performance Tests (Agilent N7840A Software Package) E8362B, E8363B, E8364B
Dynamic Accuracy Test
Function of the Test: To measure the relative power linearity of the analyzer’s receivers.
Specification Tested: Test Port Input–Dynamic Accuracy
Equipment Used: The dynamic accuracy test set (Z5623A Option H01), a power meter
(E4418B or E4419B) power sensors, and two test cables.
Description of the Test:
1. The analyzer’s test ports are tested separately at a specific CW frequency and a
reference power level of 20 dBm.
2. The analyzers driving port is connected to the dynamic accuracy test set’s source port
and the analyzers receiving port is connected to the dynamic accuracy test set’s
receiver port. A power sensor is connected to the dynamic accuracy test sets power
meter port.
3. The dynamic accuracy test set is used to input a signal from the driving port on the
analyzer. The input signal is routed through step attenuators to both the analyzer and
a power sensor.
4. The analyzer source and the dynamic accuracy test set create power levels that are
“deltas” from the reference power level of 20 dBm. Each delta is measured two ways:
by the power sensor and by the analyzers receiver under test.
5. The analyzers power level into the port under test is set to the 20 dBm reference
level.
6. The power level is measured with the power sensor: (P
mr
).
7. The power level is measured using the analyzers receiver under test: (P
pr
).
8. The power level into the analyzers port under test is stepped, in 5 dB steps, over the
range of 0 to 120 dBm. Each power level represents a “delta” from the reference level.
9. At each step, the power level is measured using the power meter: (P
md
).
10. At each step, the power level is measured using the receiver under test: (P
pd
).
11. The power error in dB = (P
mr
P
md
)(P
pr
P
pd
).

Table of Contents

Other manuals for Agilent Technologies E8364B

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies E8364B and is the answer not in the manual?

Agilent Technologies E8364B Specifications

General IconGeneral
BrandAgilent Technologies
ModelE8364B
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals