Service Guide E8364-90026 4-31
PNA Series Microwave Network Analyzers Troubleshooting
E8362B, E8363B, E8364B Measurement System Troubleshooting
Verifying the A, B, R1, and R2 Traces (Standard S-Parameter Mode)
NOTE
There is no way to view the frequency offset receiver response (Option 080).
However, some standard S-parameter receiver trace information is helpful in
troubleshooting the frequency offset section of the PNA. It is therefore
recommended that you run this test even if you only suspect the frequency
offset section of malfunctioning.
The first step is to verify that the A, B, R1, and R2 traces are present and that they are
approximately level:
•On the
System menu, point to Service, Utilities, and then click Receiver Display.
• Traces A, B, R1, and R2 are displayed in four separate data windows as shown in Figure
4-9. Identifying discrepancies of the traces in these windows can help you to isolate the
faulty assembly.
Figure 4-9 Typical Four Channel Display

Table of Contents

Other manuals for Agilent Technologies E8364B

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies E8364B and is the answer not in the manual?

Agilent Technologies E8364B Specifications

General IconGeneral
BrandAgilent Technologies
ModelE8364B
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals