3-46 Service Guide E8364-90026
Tests and Adjustments PNA Series Microwave Network Analyzers
Adjustments E8362B, E8363B, E8364B
6. Connect the power sensor 8487A and the 11901D adapter or the E4413A power sensor
to the end of the right angle adapter as shown in Figure 3-19 on page 3-43.
7. Reconnect the power and USB cables. Turn on the analyzer.
8. On the System menu, point to Service, Adjustments, and then click LO Power Adjust.
9. Ensure the GPIB settings are correct.
10. Click Begin Adj, and then follow the instructions as they are displayed.
11. Turn off the analyzer, disconnect the power cable, reconnect the semi-rigid cables, and
replace the covers.