7-12 Service Guide E8364-90038
Repair and Replacement Procedures PNA Series Microwave Network Analyzers
Removing and Replacing Front Panel Subassemblies E8362C, E8363C, E8364C
Removing and Replacing Front Panel Subassemblies
Tools Required
T-10 TORX driver (set to 9 in-lb)
T-20 TORX driver (set to 21 in-lb)
5/16 inch open-end torque wrench (set to 10 in-lb)
ESD grounding wrist strap
Refer to Figure 7-4 for the following procedures.
Pre-replacement Procedure
1. Disconnect the power cord.
2. Remove the front panel assembly. Refer to “Removing and Replacing the Front Panel Assembly” on page
7-10.
Replacing the Inverter Board
1. Remove the inverter board shield by removing the two retaining screws (item
).
2. Disconnect the two A2 display cables (item
) from the inverter board.
3. Disconnect the inverter cable (item
) from the inverter board.
4. Remove the two retaining screws (item
) from the inverter board.
5. Remove the inverter board from the front panel assembly.
6. Install the new inverter board by reversing the removal instructions.
7. Perform the post-repair adjustments, verifications, and performance tests that pertain to this
replacement procedure. Refer to Table 7-2 on page 7-82.
Replacing the Touch Screen Controller Board
1. Disconnect the touch screen display cable (item
) from the touch screen controller board.
2. Disconnect the touch screen controller cable (item
) from the touch screen controller board. NOTE THAT THE
CABLE CONNECTS TO THE UPPER FIVE PINS ON THE CONTROLLER BOARD CONNECTOR. THIS IS
IMPORTANT FOR RE-INSTALLATION.
3. Remove the two retaining screws (item
) from the touch screen controller board.
4. Remove the touch screen controller board from the front panel assembly.
5. Install the new touch screen controller board by reversing the removal instructions.
6. Perform the post-repair adjustments, verifications, and performance tests that pertain to this
replacement procedure. Refer to Table 7-2 on page 7-82.

Table of Contents

Other manuals for Agilent Technologies E8364C

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies E8364C and is the answer not in the manual?

Agilent Technologies E8364C Specifications

General IconGeneral
BrandAgilent Technologies
ModelE8364C
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals