1290 Infinity Quaternary Pump User Manual 103
Error Information
7
Pump Error Messages
Pump Error Messages
These errors are pump specific.
Pressure of quaternary pump above upper limit
Error ID: 29163
The pressure has exceeded the upper pressure limit.
Parameter: Measured pressure
Pressure below lower limit
Error ID: 29176
The pressure has dropped below the lower limit.
Parameter: None
Probable cause Suggested actions
1 Blockage in flow path after the pressure
sensor.
Check for blockages in the LC system, e.g.
purge valve, Jet Weaver, degraded column,
column frits, needle, needle seat, capillaries
etc.
Check for particles in the solvent.
2 Inappropriate settings (pressure limit, flow
rate).
Decrease flow rate.
Increase pressure limit.
Probable cause Suggested actions
1 Leak
Check for leaks.
2 Bottle empty
Check bottle filling.
3 Wrong solvent (viscosity)
Check solvent.
4 Inappropriate setting
Check flow rate and lower pressure limit.
5 Column degradation
Replace column.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies G4204A and is the answer not in the manual?

Agilent Technologies G4204A Specifications

General IconGeneral
BrandAgilent Technologies
ModelG4204A
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals