88 1290 Infinity Quaternary Pump User Manual
5 How to Optimize the Performance of Your Module
Using Solvent Calibration Tables
Using Solvent Calibration Tables
Importing Solvent Calibration Tables
RC.NET based Agilent graphical user interfaces (ChemStation, EZChrom
Elite, OpenLab etc.) include data for most commonly used solvents in
HPLC. This data contains solvent properties and is used for optimum
pump control in order to ensure best flow and composition accuracy.
If your solvent is not included to the software, please check the Agilent
web site
http://www.chem.agilent.com/_layouts/agilent/downloadFirmware.aspx?whid=69761
for additional libraries (registration required), which also provides updates
and optimized data.
If your solvent is neither available in the user interface nor in the library,
please use generic solvents. "Generic aqueous" gives good results for most
solvent mixtures with at least 50 % water, which have similar properties
as pure water. For other solvents with high organic percentage, "Generic
organic" gives a good approximation.
Importing Solvent Calibration in ChemStation
1 Go to menu Instrument > Instrument configuration.
2 In the Instrument Configuration screen choose your module and click
Configure.
3 Click Configure Solvent Type Catalogs.
4 In Solvent Type Catalogs click Import.
5 Navigate to the location of the solvent calibration table and click Open.
6 The new solvent will now appear in the Solvent Type Catalogs.
7 The imported solvent is now available for selection as a solvent type,
see Table 4 on page 63.