EasyManuals Logo
Home>Agilent Technologies>Laboratory Equipment>G4204A

Agilent Technologies G4204A User Manual

Agilent Technologies G4204A
302 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #263 background imageLoading...
Page #263 background image
1290 Infinity Quaternary Pump User Manual 263
LAN Configuration
13
Initialization Mode Selection
Initialization Mode Selection
The following initialization (init) modes are selectable:
Bootp
When the initialization mode Bootp is selected, the module tries to
download the parameters from a Bootp Server. The parameters obtained
become the active parameters immediately. They are not stored to the
non- volatile memory of the module. Therefore, the parameters are lost
with the next power cycle of the module.
Figure 39 Bootp (Principle)
Tabl e 14 Initialization Mode Switches
SW 6SW 7SW 8Init Mode
OFF OFF OFF Bootp
OFF OFF ON Bootp & Store
OFF ON OFF Using Stored
OFF ON ON Using Default
ON OFF OFF
DHCP
1
1
Requires firmware B.06.40 or above. Modules without LAN on board, see G1369C LAN Interface Card