1290 Infinity Quaternary Pump User Manual 115
Error Information
7
Pump Error Messages
Pump drive stop not found
Error ID: 29207
The pump drive stop has not been found.
• Parameter: 1 – 2 referring to pump drive
Timeout: Wait for Composition
Error ID: 29180
A target condition (composition) has been sent to the instrument which
should have been reached within an expected time frame but didn’t.
Either the limit, time frame or the current value of the variable has been
modified later directly or indirectly.
Probable cause Suggested actions
1 Pump drive spindle is defect.
Please contact your Agilent service
representative.
Probable cause Suggested actions
1 Incorrect parameters have been sent to the
instrument by the control software or
manual changes.
Verify control software, macros, manual
commands.