1290 Infinity Quaternary Pump User Manual 261
LAN Configuration
13
TCP/IP Parameter Configuration
TCP/IP Parameter Configuration
To operate properly in a network environment, the LAN interface must be
configured with valid TCP/IP network parameters. These parameters are:
IP address
Subnet Mask
Default Gateway
The TCP/IP parameters can be configured by the following methods:
by automatically requesting the parameters from a network- based
BOOTP Server (using the so- called Bootstrap Protocol)
by automatically requesting the parameters from a network- based DHCP
Server (using the so- called Dynamic Host Configuration Protocol). This
mode requires a LAN- onboard Module or a G1369C LAN Interface card,
see “Setup (DHCP)” on page 268
by manually setting the parameters using Telnet
by manually setting the parameters using the Instant Pilot (G4208A)
The LAN interface differentiates between several initialization modes. The
initialization mode (short form ‘init mode’) defines how to determine the
active TCP/IP parameters after power- on. The parameters may be derived
from a Bootp cycle, non- volatile memory or initialized with known default
values. The initialization mode is selected by the configuration switch, see
Table 14 on page 263.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies G4204A and is the answer not in the manual?

Agilent Technologies G4204A Specifications

General IconGeneral
BrandAgilent Technologies
ModelG4204A
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals