1290 Infinity Quaternary Pump User Manual 99
Error Information
7
General Error Messages
Remote Timeout
Error ID: 0070
A not- ready condition is still present on the remote input. When an
analysis is started, the system expects all not- ready conditions (for
example, a not- ready condition during detector balance) to switch to run
conditions within one minute of starting the analysis. If a not- ready
condition is still present on the remote line after one minute the error
message is generated.
Lost CAN Partner
Error ID: 0071
During an analysis, the internal synchronization or communication
between one or more of the modules in the system has failed.
The system processors continually monitor the system configuration. If one
or more of the modules is no longer recognized as being connected to the
system, the error message is generated.
Probable cause Suggested actions
1 Not-ready condition in one of the
instruments connected to the remote line.
Ensure the instrument showing the not-ready
condition is installed correctly, and is set up
correctly for analysis.
2 Defective remote cable.
Exchange the remote cable.
3 Defective components in the instrument
showing the not-ready condition.
Check the instrument for defects (refer to the
instrument’s documentation).
Probable cause Suggested actions
1 CAN cable disconnected.
• Ensure all the CAN cables are connected
correctly.
• Ensure all CAN cables are installed
correctly.
2 Defective CAN cable.
Exchange the CAN cable.
3 Defective main board in another module.
Switch off the system. Restart the system, and
determine which module or modules are not
recognized by the system.