286 1290 Infinity Quaternary Pump User Manual
13 LAN Configuration
Manual Configuration
With the Instant Pilot (G4208A)
To configure the TCP/IP parameters before connecting the module to the
network, the Instant Pilot (G4208A) can be used.
1 From the Welcome screen press the More button.
2 Select Configure.
3 Press the DAD button.
4 Scroll down to the LAN settings.
Figure 57 Instant Pilot - LAN Configuration
5 Press the Edit button (only visible if not in Edit mode), perform the
required changes and press the Done button.
6 Leave the screen by clicking Exit.