70 1290 Infinity Quaternary Pump User Manual
4 Using the Pump
Flushing the Filter
Flushing the Filter
For highest performance and robustness, the 1290 Infinity Quaternary
Pump uses 3 solvent filters:
1 Solvent inlet filter, 20 µm pore size (5041- 2168) as part of Bottle Head
Assembly (G4220- 60007) have a large pore size of about 20 µm and
filter out particles before they reach the pump.
2 An outlet filter (average pore size 5 µm; Outlet filter 1290 Infinity
Quaternary Pump (G4204- 60004)) between pump head and pressure
sensor filters out particles which may be created in the pump by wear
of piston or wash seals.
This filter can be replaced as required.
Solvent inlet filter