1290 Infinity Quaternary Pump User Manual 117
Error Information
7
Pump Error Messages
Timeout: Wait for Flow
Error ID: 29183
A target condition (flow rate) has been sent to the instrument which
should have been reached within an expected time frame but didn’t.
Either the limit, time frame or the current value of the variable has been
modified later directly or indirectly.
Timeout: Wait for Pressure
Error ID: 29185
A target condition (pressure) has been sent to the instrument which
should have been reached within an expected time frame but didn’t.
Either the limit, time frame or the current value of the variable has been
modified later directly or indirectly.
Probable cause Suggested actions
1 Incorrect parameters have been sent to the
instrument by the control software or
manual changes.
Verify control software, macros, manual
commands.
Probable cause Suggested actions
1 Incorrect parameters have been sent to the
instrument by the control software or
manual changes.
Verify control software, macros, manual
commands.
2 Leak
Run system pressure test for identifying and
localizing the leak. Tighten leak.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies G4204A and is the answer not in the manual?

Agilent Technologies G4204A Specifications

General IconGeneral
BrandAgilent Technologies
ModelG4204A
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals