132 1290 Infinity Quaternary Pump User Manual
9 Maintenance
Replacing the Pressure Sensor
1
Remove capillary connections between the pressure
sensor and the Multi Purpose Valve, and between the
pressure sensor and the outlet filter of the secondary
pump head, respectively.
2 Remove the screws that fix the pressure sensor to the
chassis.
3 Carefully pull out the pressure sensor for about 2 cm.
Then unscrew the cable from the pressure sensor.
4 Connect the new pressure sensor to the pressure sensor
connector.