190 1290 Infinity Quaternary Pump User Manual
9 Maintenance
Replacing the Outlet Filter
3
Mount the new outlet filter. Using a torque wrench set to
16 Nm is recommended.
4 Mount the capillary connection to the pressure sensor.
16 Nm