5 Operating in EI Mode
To change the network settings of the MSD
110 5977B Series MSD Operating Manual
6 Confirm that BootP is set to OFF. If your LAN assigns IP addresses using a
BootP server, click ON and skip the next step.
7 To update the MSD IP address, Gateway IPA, and SubNet Mask enter the
new values. Before clicking submit, you can recover the previous settings by
clicking Return to Main Menu and then returning here.
8 Click Submit to upload this net configuration to the MSD.
A dialog opens to confirm that the network configuration process has
started. (See Figure 30.)
9 Click OK to close the dialog and wait for the prompt requesting to Manually
reboot of the MSD/SmartCard to activate the new Settings.
10 Use the MSD on/off switch to reboot the MSD SmartCard.
Figure 30 IP address configuration