7 Error Information
General Error Messages
1260 Infinity II Flexible Pump User Manual 122
General Error Messages
General error messages are generic to all Agilent series HPLC modules and may
show up on other modules as well.
Timeout
Error ID: 0062
The timeout threshold was exceeded.
Probable cause Suggested actions
1 The analysis was completed successfully,
and the timeout function switched off the
module as requested.
Check the logbook for the occurrence and source
of a not-ready condition. Restart the analysis
where required.
2 A not-ready condition was present during a
sequence or multiple-injection run for a
period longer than the timeout threshold.
Check the logbook for the occurrence and source
of a not-ready condition. Restart the analysis
where required.