1260 Infinity II Flexible Pump User Manual 123
7 Error Information
General Error Messages
Shutdown
Error ID: 0063
An external instrument has generated a shutdown signal on the remote line.
The module continually monitors the remote input connectors for status signals.
A LOW signal input on pin 4 of the remote connector generates the error
message.
Probable cause Suggested actions
1 Leak detected in another module with a CAN
connection to the system.
Fix the leak in the external instrument before
restarting the module.
2 Leak detected in an external instrument with
a remote connection to the system.
Fix the leak in the external instrument before
restarting the module.
3 Shut-down in an external instrument with a
remote connection to the system.
Check external instruments for a shut-down
condition.
4 The degasser failed to generate sufficient
vacuum for solvent degassing.
Check the vacuum degasser for an error
condition. Refer to the Service Manual for the
degasser or the pump that has the degasser
built-in.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies G7104A and is the answer not in the manual?

Agilent Technologies G7104A Specifications

General IconGeneral
BrandAgilent Technologies
ModelG7104A
CategoryWater Pump
LanguageEnglish

Related product manuals