1260 Infinity II Flexible Pump User Manual 143
7 Error Information
Pump Error Messages
Timeout: Wait for Composition
Error ID: 29180
A target condition (composition) has been sent to the instrument which should
have been reached within an expected time frame but didn’t. Either the limit, time
frame or the current value of the variable has been modified later directly or
indirectly.
Timeout: Wait for run volume
Error ID: 29181
A target condition (run volume, which is the volume delivered since the method
run start) has been sent to the instrument which should have been reached
within an expected time frame but didn’t. Either the limit, time frame or the
current value of the variable has been modified later directly or indirectly (for
example the flow rate).
Probable cause Suggested actions
1 Incorrect parameters have been sent to the
instrument by the control software or
manual changes.
Verify control software, macros, manual
commands.
Probable cause Suggested actions
1 Incorrect parameters have been sent to the
instrument by the control software or
manual changes.
Verify control software, macros, manual
commands.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies G7104A and is the answer not in the manual?

Agilent Technologies G7104A Specifications

General IconGeneral
BrandAgilent Technologies
ModelG7104A
CategoryWater Pump
LanguageEnglish

Related product manuals