7 Error Information
General Error Messages
Fan Failed
Error ID: 0068
The cooling fan in the module has failed.
The hall sensor on the fan shaft is used by the mainboard to monitor the fan
speed. If the fan speed falls below a certain limit for a certain length of time, the
error message is generated.
Depending on the module, assemblies (e.g. the lamp in the detector) are turned
off to assure that the module does not overheat inside.
Leak
Error ID: 0064
A leak was detected in the module.
The signals from the two temperature sensors (leak sensor and board-mounted
temperature-compensation sensor) are used by the leak algorithm to determine
whether a leak is present. When a leak occurs, the leak sensor is cooled by the
solvent. This changes the resistance of the leak sensor which is sensed by the
leak sensor circuit on the main board.
Probable cause Suggested actions
1 Fan cable disconnected.
Please contact your Agilent service representa-
tive.
2 Defective fan.
Please contact your Agilent service representa-
tive.
3 Defective mainboard.
Please contact your Agilent service representa-
tive.
Probable cause Suggested actions
1 Loose fittings.
Ensure all fittings are tight.
2 Broken capillary.
Exchange defective capillaries.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies G7117B and is the answer not in the manual?

Agilent Technologies G7117B Specifications

General IconGeneral
BrandAgilent Technologies
ModelG7117B
CategorySecurity Sensors
LanguageEnglish

Related product manuals