EasyManuals Logo

Agilent Technologies G7117B User Manual

Agilent Technologies G7117B
230 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #131 background imageLoading...
Page #131 background image
Agilent InfinityLab LC Series Diode Array Detectors User Manual 131
8 Test Functions and Calibration
Intensity Test
Figure 36 Intensity Test – Signals
Test Failed
Intensity Test Evaluation
Probable cause Suggested actions
1 Absorbing solvent or air bubble in flow cell.
Ensure the flow cell is filled with water, and
free from air bubbles.
Repeat test with Max-Light Cartridge Test
Cell and compare results.
2 Incorrect calibration
Recalibrate and repeat the test.
3 Dirty or contaminated flow cell.
Run the cell test. If the test fails, flush the flow
cell. See also “Clean the Max-Light Cartridge
Cell” on page 165.
4 Dirty or contaminated optical components.
Please contact your Agilent service representa-
tive.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies G7117B and is the answer not in the manual?

Agilent Technologies G7117B Specifications

General IconGeneral
BrandAgilent Technologies
ModelG7117B
CategorySecurity Sensors
LanguageEnglish

Related product manuals