Agilent InfinityLab LC Series Diode Array Detectors User Manual 133
8 Test Functions and Calibration
Cell Test
1 Run the Cell-Test with Agilent Lab Advisor (for further information see
Online-Help of user interface).
Figure 37 Cell Test – Results
Figure 38 Cell Test – Signals (example shows low intensity for flow cell)
Test Failed (low ratio value)
Cell Test Evaluation
Probable cause Suggested actions
1 Absorbing solvent or air bubble in flow cell.
Ensure the flow cell is filled with water, and free
from air bubbles.
2 Dirty or contaminated flow cell.
Clean the flow cell as described in “Clean the
Max-Light Cartridge Cell” on page 165.