Agilent InfinityLab LC Series Diode Array Detectors User Manual 137
8 Test Functions and Calibration
ASTM Drift and Noise Test
ASTM Drift and Noise Test
The ASTM noise test determines the detector noise over a period of 20 minutes.
The test is done with installed Max-Light Cartridge Cell or Max-Light Cartridge
Test Cell.
This test does also check for the drift. It is also part of the “Self Test” (without
checking for the drift).
If the test is performed with the Max-Light Cartridge Test Cell, the test results are
not influenced by solvent or pump effects.
When
In case of noise and drift problem.
Parts required # Description
1 Max-Light Cartridge Cell (filled with water)
OR 1 Max-Light Cartridge Test Cell
Preparations
Detector and UV-lamp must be on for at least 2 hours.
ASTM measurements based on specifications may require longer stabilization times.
When using a Max-Light Cartridge Cell a flow rate of 0.5 mL/min with water is required.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies G7117B and is the answer not in the manual?

Agilent Technologies G7117B Specifications

General IconGeneral
BrandAgilent Technologies
ModelG7117B
CategorySecurity Sensors
LanguageEnglish

Related product manuals