Agilent InfinityLab LC Series Diode Array Detectors User Manual 167
9 Maintenance
Correcting Leaks
Correcting Leaks
When
If leak has occurred.
Tools required Description
Tissue
Preparations
Turn the pump off.
1 Open the doors. 2 Use tissue to dry the leak sensor area (1). Observe the
capillary connections and the flow cell area (2) for leaks
and correct, if required.
3 Close the doors.