Agilent InfinityLab LC Series Diode Array Detectors User Manual 79
5 Optimizing the Detector
Optimization Overview
High Dynamic Range
The Max-Light Cartridge HDR cell can be used with the G7117A DAD FS and
G7117B DAD. The cell is required as a part of the High Dynamic Range (HDR)
solution.
Recommendations
For G4212-60007 and G4212-60008
For G4212-60007 and G4212-60008
The use of Peek-FS capillaries is not recommended. In combination with the SST
zero dead volume fitting (e.g. at the inlet) the capillary could break and the glass
particles could block/damage the flow cell.
NOTE
To protect the flow cell against overpressure (e.g. in systems with LC/MS) install
Inline Pressure Relief Valve Kit (G4212-68001), see “Inline Pressure Relief Valve
Kit (G4212-68001)” on page 81.