Replacement Parts
Description Part No.
MS interface column nut, female 05988-20066
Column nut for long or long two-hole ferrules 05921-21170
Universal column nut, 2/pk 5181-8830
Remedy
Check interface nut for tightness. Replace if necessary.
Check and leak test the GC injection port.
Symptoms of leaks
Higher than normal vacuum manifold pressure or foreline pressure
Higher than normal background
Peaks characteristic of air (m/z 18, 28, 32, and 44 or m/z 14 and 16)
Poor sensitivity
Low relative abundance of m/z 502 (this varies with the tune program and MSD used)
MS interface column nut, 05988-20066
Universal column nut, 5181-8830