Syringe Characteristics and Recommended Uses
Syringe Advantage Limitations Recommended Use
10 µL, PTFE-tipped
Less plunger binding than fitted
plunger
Replaceable plunger for reduced
repair cost
Tight seal between plunger and barrel
More expensive than fitted plunger
PTFE-tipped syringes not available
in 5 µL size
High sample throughput
Samples in polar solvents
Dirty samples
Gases and volatile samples
Reactive samples
10 µL, fitted plunger
Most economical
Most reliable fitted plunger syringe
Less bending
Better for high viscosity samples
Most accurate only for 1 µL and
larger injections
Plunger not replaceable
General purpose syringe
Clean samples
Routine analysis
5 µL, fitted plunger
Most accurate for 1 µL injection
No hardware modification needed for
0.5 µL
Thinnest plunger, can bend more
easily
Not ideal for higher viscosity
samples
Plunger not replaceable
1 µL injections
Clean samples
Routine analysis

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies gc/ms series and is the answer not in the manual?

Agilent Technologies gc/ms series Specifications

General IconGeneral
BrandAgilent Technologies
Modelgc/ms series
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals