Gas Management
Agilent recommends that carrier
and detector gases be 99.9995% pure.
Air needs to be zero grade or better.
Agilent also recommends using traps to
remove hydrocarbons, water and oxygen.
When used with capillary columns, GC
detectors require a separate makeup gas
for optimum sensitivity. For each detector
and carrier gas, there is a preferred choice
for makeup gas. This table lists gas
recommendations for capillary columns.
Detector Carrier Gas Make up
1st choice
Make up
2nd choice
Purge or
Reference
Electron capture Hydrogen Argon/methane Nitrogen Anode purge must
be same as
makeup
Helium Argon/methane Nitrogen
Nitrogen Nitrogen Argon/methane
Argon/methane Argon/methane Nitrogen
Flame ionization Hydrogen Nitrogen Helium Hydrogen and air
for detector
Helium Nitrogen Helium
Nitrogen Nitrogen Helium
Flame
photometric
Hydrogen Nitrogen None Hydrogen and air
for detector
Helium Nitrogen
Nitrogen Nitrogen
Argon Nitrogen
Mass selective Hydrogen None None
Helium None None
Nitrogen
phosphorous
Helium Nitrogen Helium Hydrogen and air
for detector
Nitrogen Nitrogen Helium
Thermal
conductivity
Hydrogen Must be same as
carrier and
reference
Must be same as
carrier and
reference
Reference must
be same as carrier
and makeup
Helium
Nitrogen
Gas Purity and Selection