Site Preparation
Assembling the gas plumbing
17
Making swage-type connections
Assemble fitting,
bottom tubing
Hand-tighten,
then wrench-tighten
Swage-Type Fittings
Assemble the fittings as shown. Push the tubing into the fitting as far as
it will go, then tighten the fitting by hand (no wrench). Use a wrench to
tighten further (3/4•turnfor 1/8•inchtubing, 1•1/4•turnfor 1/4•inch
tubing).