EasyManua.ls Logo

Agilent Technologies U1231A - Calibration Process

Agilent Technologies U1231A
55 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
1 Calibration Procedures
Calibration Process
6 U1231A/U1232A/U1233A Service Guide
Calibration Process
1 Prior to performing the verification tests, see the “Test
Considerations” on page 7.
2 Perform the verification tests to characterize the
multimeter; see “Performance Verification Tests” on
page 8.
3 Unsecure the multimeter for calibration; see “Calibration
Security” on page 12.
4 Prior to performing the adjustments, see the “Adjustment
considerations” on page 18.
5 Perform the adjustment procedure; see “Adjustment
procedure” on page 19.
6 Secure the multimeter against unauthorized calibration;
see “Exiting the adjustment mode” on page 26. Ensure
that the multimeter has quit the adjustment mode and is
turned off.
7 Record the new security code and calibration count in the
multimeter's maintenance records.

Table of Contents

Other manuals for Agilent Technologies U1231A

Related product manuals