Calibration Procedures 1
Test Considerations
U1231A/U1232A/U1233A Service Guide 7
Test Considerations
For optimum performance, all procedures should comply
with the following recommendations:
The performance verification test or adjustment should be
performed under a laboratory condition, where the
ambient temperature can be controlled.
Ensure that the calibration ambient temperature is stable
and is between 18 °C and 28 °C. Ideally the calibration
should be performed at 23 °C ± 1 °C.
Ensure that the ambient relative humidity is less than
80%.
The instrument should be in the laboratory environment
for at least 1 hour prior.
Allow a warm- up period of 3 minutes.
Use shielded twisted- pair PTFE- insulated cables to reduce
settling and noise errors. Keep the input cables as short
as possible.