EasyManua.ls Logo

Aidcall Touchsafe Pro - Zone Routing Setup; Common Calls Configuration; Telecare Device Setup

Aidcall Touchsafe Pro
33 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Doc No. FM5600 issue E Page 11
10. INTERFACE INSTALLATION
The Interface requires power from the 12V/1A battery backed PSU and RS232 data from the Pager Port on
any Touchsafe Pro Nursecall panel.
The Interface unit is normally located in close proximity to the Nursecall panel and is connected using the 2
metre Patch Lead provided. Note: this cable can be extended up to 20 metres maximum if necessary.
The Interface converts RS232 data from the panel to RS485 for connection to the Master and Slave Bases.
Note: the maximum length for the RS485 bus cable is 600 metres.
INTERFACE FIXING
Undo the single screw fixing in the centre and remove the Interface lid.
Choose a location to fix the Interface within 2 metres cable run of a Nursecall panel.
Fix the Interface to the wall using the screws provided through the 2 fixing holesA” shown below.
PATCH LEAD
Cut an RJ45 plug off one end of the 2 metre Patch Lead (Aidcall Part No. W00402).
Strip-back the cable and identify the green, orange and white/blue wires.
30mm
60mm
A
A
Undo Fixing
Screw and
Remove Lid
Cut plug off one end
W00402 - 2m Patch Lead
Strip-back 3 wires
as shown here
RS232 from Panel
RS485 & Power

Other manuals for Aidcall Touchsafe Pro