B1: Trouble Shooting
Fault
Possible Explanation
Action
Notes
Unit does not turn
on, screen remains
blank and LCD back
light is not
illuminated
Battery low level
Connect the supplied
power supply unit for
a minimum of 12
hours charging to
achieve full charge
Unit fails AutoCal
with Sensor Open
Circuit Error
displayed or
Instrument
indicates lower
limit
It may be that an
internal wiring fault
has occurred
Perform AutoCal again,
ensuring that the
desiccant head is fully
extended
If action fails
contact your
authorised
distributor
Instrument will not
dry down to lower
limit
Desiccant Sieve wet
Perform service on unit
using Service Kit
Service Kit
available from
your authorised
distributor
Sealing ‘O’ ring failure
Perform service on unit
using Service Kit
Battery has short
life between
charges
Battery reaching end
of useful life
Contact your
authorised distributor
Navigation Keypad
fails to react to
user key presses
The keypad may be
damaged or an
internal wiring fault
has occurred
Contact your
authorised distributor
Fails to connect PC
on USB Channel
Cable damaged or
cable too long.
Change Cable
General Error in operation
Send a SADPmini2 Report to your
authorised distributor
(See Appendix D5)

Table of Contents

Other manuals for Alpha Moisture Systems SADPmini2

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Alpha Moisture Systems SADPmini2 and is the answer not in the manual?

Alpha Moisture Systems SADPmini2 Specifications

General IconGeneral
BrandAlpha Moisture Systems
ModelSADPmini2
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals