54
D7: Downloading Calibration Files
Calibration files may be distributed as email attachments when a new sensor has been
installed on the instrument. Users should save the file (.bin) from the email to a
location on their PC before updating their instrument.
To install the calibration file:
1. Select Calibration ① from the Headings Menu.
2. Select Browse ②.
3. Locate and select the required Calibration file on your computer. NOTE: Calibration
files are Binary files (.bin).
4. Select a file to load. NOTE: The calibration file details will be displayed in the dialogue
box ③.
5. Select Download ④ to transfer the file to the SADPmin2.
6. Select Exit ④.