34 | 5000 Single-Point Moisture Analyzer
Computer Operating Properly Alarm
This alarm, also known as the COP or watchdog alarm, activates when normal CPU logic
flow is disrupted. Possible triggers of this alarm are:
power failure
internal power failure
software problems, or
extreme electronic interference.
When activated, the COP alarm generates a system reset but retains software configura-
tion parameters. You do not have to reconfigure the system.