Universal IV - Level Measurement System
14
2.4 Save/Print Entries
• Allvaluesdevelopedinthisconfiguration/calibration
procedure may be saved to be reloaded into another
transmitter or printed as a hard copy
• PopupscreenscomefromselectionintheFILEpull
downatthetopleftofthePCmenu
• Copiesaresavedinboth.UniversalIVCMfileand.txt
files
• The.UniversalIVCMfilewilldownloadintoa
transmitterthroughtheOPENcommandwhile.txtfiles
canbeprintedusingPRINTcommand
Figure 2-4
Print Pop-up from Menu
AMETEK Drexelbrook
205 Keith Valley Road
Horsham, PA 19044
Telephone: 215-674-1234
FAX: 215-674-2731
Service: 800-527-6297
654321 :rebmuN laireSOMED-TL :DI-gaT
Scratch Pad: DEMO CALIBRATION Software Version: 6.0
4 :noitisoP egnaR.ces 0 :emiT gnipmaD
stniopdnE egnaRnoitarugifnoC tnemurtsnI
retaW % 00.0 :)Am 4( VRL4 egnaR liO yvaeH %08 oT 0 - M :evruC tuptuO/tupnI
URV (20 mA): 80.00 % Water
Real Time View
Percent Water: 55.00 %
Capacitance: 600.00 pF
Loop Current: 15.00 mA
Percentage: 68.75 %
Status:
OK
Input/Output Table
Number of Points: 15
Input Output
pF % Water
00.000.04
00.0102.65
00.3200.97
00.5302.221
00.0400.981
00.2400.652
00.0500.593
00.5500.006
00.7500.057
00.0600.009
00.5600.0011
00.0700.0042
00.5700.0072
00.0800.0092
00.3800.0543