EasyManuals Logo

Ametek Drexelbrook Universal IV User Manual

Ametek Drexelbrook Universal IV
26 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #19 background imageLoading...
Page #19 background image
Quick Start Guide
15
2.5 Calibration / Configuration Via Display / Keypad
• AllfunctioninthePCsoftwareexceptSave/Printre
accessible via the display/keypad
• ToentertheConfigurationMenufollowthesesteps
• PressandholdtheENTERbuttonforapproximately5
seconds
• UsetheUPandDOWNbuttonstoscrollthroughthe
available selections
• PressENTERtoaccesssub-menuitems
• UseUPandDOWNbuttonstoadjustsettings-settings
thatcanbeadjustedwillbeflashing
• PressENTERtoaccepttheadjustmentorpressand
holdENTERfor5secondstoexittopreviousmenu
Menu Function
(display
abbreviation)
"Values
(display abbreviation)"
Description
"Fct 1.00
Water Cut Ranges
(RANGE)"
Select the water cut range for optimal
measurement
0 to 1% water in Light Oil (LIGHT A) 'Light Oil' is defined as oil with API Gravity
less than 25. Heavy Oil is defined as oil with
API Gravity greater than 25.
0 to 5% water in Light Oil (LIGHT B)
0 to 10% water in Light Oil (LIGHT C)
0 to 30% water in Light Oil (LIGHT D)
0 to 50% water in Light Oil (LIGHT E)
0 to 1% water in Heavy Oil (HEAVY F)
0 to 5% water in Heavy Oil (HEAVY G)
0 to 10% water in Heavy Oil (HEAVY H)
0 to 30% water in Heavy Oil (HEAVY I)
0 to 50% water in Heavy Oil (HEAVY J)
0 to 80% water in Heavy Oil (HEAVY M) -
Default
CUSTOM Custom range requires a custom strapping
table. See Fct 3.00
"Fct 2.00
Calibration (CAL)"
Enter this menu to calibrate the unit
"Fct 2.01
Indicated Calibration Point
(IND CAL)"
% water - 0.0 Default Enter the water cut reading captured at the
time of taking the sample for calibration
measurement
"Fct 2.02
Actual Calibration Point
(ACT CAL)"
% water - 0.0 Default Enter the actual water cut reading verified by
another method of water cut measurement
"Fct 3.00
Strapping table
(STRAP)"
Use this strapping table menu to define
a custom range if selected in FCT 1.00.
Otherwise the correct strapping table is
automatically loaded when range is selected in
FCT 1.00. Default values are for 'HEAVY M'
range
"Fct 3.01
Maximum points
(MAX PNT)"
15 (Default) Enter the total number of points in the
strapping table which is range dependent
"Fct 3.02
Point number index
(INDEX)"
"1..MAX PNT
1 (Default)"
Enter the point number index

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Ametek Drexelbrook Universal IV and is the answer not in the manual?

Ametek Drexelbrook Universal IV Specifications

General IconGeneral
BrandAmetek
ModelDrexelbrook Universal IV
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals