Quick Start Guide
19
3.1.5 Commissioning
Start-up checklist
Do not connect power until you have gone through the checklist
below
1.Arethewettedcomponents(gasket,flangeandsensing
element)resistanttothecorrosivepropertiesofthetank
product?
2. Does the information given on the nameplate correspond with
the application?
3.Exdapplications:Haveyouconnectedtheequipotential
bonding system correctly?
4.Exiapplications:Areyouusinganintrinsicbarrierwithin
the correct parameters?
5. Did you install cable entries of the correct internal diameter
so that there is a good seal around the cable? Are the
cable glands suitably certified per the application and the
hazardous area parameters?
6.Donotusetheearthterminalinthewiringcompartment:use
the equipotential bonding system.
3.2 The Compartment Cover
ViewportCleaning:TheviewportismadeofBorosilicateglass
and can be cleaned with any common glass cleaning product
(e.g.:Windex™,Isopropylalcohol,etc.)thatissuitableforthe
Class and Division rating of the specific system installation.
3.2.1 Opening the cover
Procedure
1. Unscrew cover stop, if applicable
2. Unscrew terminal compartment cover
3.2.2 Closing the cover
Warning: Ex d [ia] applications
Check that the terminal compartment cover is screwed tight and
thecoverstop(ifapplicable)isfastenedtightlytothecover.

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Ametek Drexelbrook Universal IV and is the answer not in the manual?

Ametek Drexelbrook Universal IV Specifications

General IconGeneral
BrandAmetek
ModelDrexelbrook Universal IV
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals